HICO-VARIOTHERM 555
HICO-VARIOTHERM 555 is the allrounder among the hypo-/hyperthermia systems and appreciated for its reliable warming and cooling of patients in intensive care units and operating rooms. Featuring a water temperature range of 5 to 40 °C, this system covers all areas of application.
Properties
- Easy and quick start-up of the non-invasive temperature control system
- Optimum temperature transfer via reusable HICO water pads
- Connections for two HICO water pads
- Optional use of Soft-Temp® single-use pads
- Nursing staff has free access to the patients since they are positioned on the water pads.
- Defibrillation-proof device
- Microprocessor-controlled Peltier technology ensures smooth and quiet operation of the powerful system.
- Compact stationary unit, available with a tilt-resistant castor stand
Areas of application
- Anesthesia
- Surgery
- Intensive-care medicine
- Cardiac / thorax surgery
- Neurosurgery
- Neurology
- Emergency medicine
- Oncology
- Orthopedics
- Rheumatology
- Pain Centers
- Stroke Units
- Toxicology
Technical Data
REF |
600001 (220 V/230 V) / 600021 (110 V/115 V) |
Nominal voltage |
220 - 230 VAC 50 Hz/60 Hz / 110 - 115 VAC 50 Hz/60 Hz |
Power input |
440 W |
Current consumption |
approx. 1.9 A (220 V/230 V) / 3.8 A (110 V/115 V) |
Nominal temperature range |
5 °C – 40 °C |
Safety shutdown |
approx. 4 °C and 41.1 - 41,5 °C (independent safety shutdown, resulting in a max. pad surface temperature of < 41 °C) |
Correction value |
0.5 °C (water temperature – temperature display) |
Pump capacity |
max. 7 l/min, max. 0.36 bar |
Warm-up time |
approx. 5 – 10 min (20 - 37 °C without water pad attached) |
Cool-down time |
approx. 5 – 10 min (20 - 10 °C without water pad attached) |
Fuse rating |
2 x T 3,15 A; L 250 V (220 V/230 V) / 2 x T 6,3 A; L 250 V (110 V/115 V) |
Class / type of protection (IEC 60601-1) |
I / BF (defibrillation-proof) |
IP protection class (IEC 60529) |
IP X1 (drip-proof) |
Risk class (Regulation (EU) 2017/745) |
II b |
Ambient temperature |
10 – 40 °C |
Relative humidity |
approx. 30 – 70 % |
Transport / storage temperature |
3 – 60 °C |
Air pressure |
700 – 1060 mbar |
Tank volume |
approx. 1.0/2.0 l (min./max.) |
Permissible height difference |
max. 1 m (between unit and water pad) |
Dimensions W x H x D |
approx. 20 x 60 x 51 cm |
Packaging dimensions |
295 x 795 x 590 mm |
Weight |
approx. 25 kg (empty water tank) |
Noise emission |
approx. 45 dB(A) |
Alarm level |
> 55 dB(A) |
Tank volume heat exchanger |
1.5 l |
Coolant |
water-glycol mixture |
Test basis |
Regulation (EU) 2017/745: NB MDR 0598, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-35 |
EMDN code |
Z12040208 - Body thermoregulation equipment |
Accessories
Water pads (multi-use)
- Uniform distribution of temperature across the entire pad surface thanks to the unique honeycomb structure
- Radiolucent, lightweight, durable and elastic
- Free of latex and phthalates
- Easy cleaning of the multi-use pads by standard wipe disinfection
- High operational safety as HICO water pads are free of electrical connections
Water pad - Cover |
60 x 70 cm |
REF 550025 |
Water pad - Cover |
120 x 70 cm |
REF 550026 |
Water pad - Intensive |
170 x 50 cm |
REF 550046 |
Water pad - Intensive |
92 x 50 cm |
REF 550047 |
Water pad - Surgery |
170 x 35 cm |
REF 550048 |
Water pad - Surgery |
92 x 35 cm |
REF 550049 |
Water pad - Baby |
50 x 30 cm |
REF 550044 |
Castor stands and brackets
Tilt-resistant safety stand (4-castors) | REF 602810 |
Hose extensions
Hose extension (3 m), incl. quick-release couplings, for connecting device and single-use pads | REF 550069 |
Insulated hose extension (2 m), incl. quick-release couplings, for connecting device and HICO water pad | REF 600022 |
Disinfection
Sanosil S003 – Disinfectant solution consisting of two basic ingredients (hydrogen peroxide and silver) for use as water additive in all HICO devices and water pads | REF 830111 |
Sanosil test strips – Quick determination of the Sanosil disinfectant concentration in liquids. 100 strips/sales unit | REF 830112 |